<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\NVSStyleSheet.xsl"?>
<Parameter>
<ParameterFamily>head</ParameterFamily>
<LongName>
diag0
</LongName>
<Name>
diag0
</Name>
<AppliesTo>
<UnnumberedList>
<ListItem><TextApplies>Thermal</TextApplies></ListItem>
</UnnumberedList>
</AppliesTo>
<Description>
<DescriptionText><text>
The first four bits specify the signal sent to pin 1 of the HC board test point
cable accessed via the PD board.
</text></DescriptionText>
<DescriptionText><text>
Setting Signal
</text></DescriptionText>
<DescriptionText><text>
0 focus error
1 lens positioning servo error
2 write laser power servo error
3 beam positioning servo error
</text></DescriptionText>
<DescriptionText><text>
The second four bits specify data stored to the lower half of the second bank
when diag5 &gt; 0.  Note that in order to store focus data, diag1 must be set to
0 as well.
</text></DescriptionText>
<DescriptionText><text>
Storing Data
</text></DescriptionText>
<DescriptionText><text>
0	focus data (stored in swaths)
16	write laser (M2) sensor
32	BP0 sensor
48	BP1 sensor
64	nosepiece sensor
80	diode temperature
96	ambient temperature
112	case temperature
128	humidity
144	LV temperature
160	focus sum
176	focus difference
192	diagnostic port 0 output
208	diagnostic port 1 output
224	focus lens position
</text></DescriptionText>
</Description>
<Type>
integer
</Type>
<Default>
<UnnumberedList>
<li>0</li>
</UnnumberedList>
</Default>
<Range>
<UnnumberedList>
<li>0 to 255</li>
</UnnumberedList>
</Range>
<ArraySize>
</ArraySize>
<Notes>
<DescriptionText><text>
</text></DescriptionText>
</Notes>
<Partition>
cfg
</Partition>
<SeeAlso>
<UnnumberedList>
<li><Link><LinkAddress>../nvs/head_cfg_diag1.xml</LinkAddress><LinkTitle>NVS: head diag1</LinkTitle></Link></li>
<li><Link><LinkAddress>../nvs/head_cfg_diag5.xml</LinkAddress><LinkTitle>NVS: head diag5</LinkTitle></Link></li>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<li><Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link></li>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:37:04 2008 PST
</Updated>
<Links>
</Links>
</Parameter>